Beamline>>Beamline

Beamline

The BioNSRRC has currently four beamlines (BL13C1, BL13B1, BL15A1,  and TPS 05A) in operation. BL13B1, BL15A1, and TPS 05A are capable of performing MAD experiments, as well as standard monochromatic crystallography. BL13C1 is designed to operate at fixed energy above the Se K edge. All four beamlines also serve as the high-throughput Synchrotron Protein Crystallography Facility (SPXF) of the National Core Facility Program for Biotechnology (NCFPB) to provide the NCFPB users with reliable, rapid and fully automated X-ray beamlines.

BioNSRRC Beamline
  Location
TLS, NSRRC
TLS, NSRRC
TLS, NSRRC
TPS, NSRRC
  X-ray Source
Wiggler on
1.5GeV TLS
Wiggler on
1.5GeV TLS
Wiggler on
1.5GeV TLS
Undulator on
3GeV TPS
  General User Time
80%
90%
80%
75%
  Status
Operational
Operational
Operational
Operational
  Experiment
Mono
MAD
MAD
MAD/Microbeam
  Wavelength (Å)
0.975
1.77-0.69
2.21-0.8
2.26-0.62
  Energy Range (keV)
12.7
7-18
5.6-15.5
5.7-20.0
  Aperture Size (μmΦ)
200
200/150/100/75/50
200/150/100/75/50
50/30/20/10/5
  Flux After Aperture(photons/sec)
3.5x1010
12/7/3/2/0.8x1010
14/9/4/2/1.1x10 10
270/140/63/38/8.2x1010
  Flux Density (photons/sec/μm2)
3.3x10 6
3.3–4.8x10 6
5.7–7.5x10 6
2.6–3.3x10 9
  Time to Henderson Limit (sec)
12121
8333–12121
5333–7018
12.1–15.4
  Detector 
Q315r
Q315r
MX300HE
MX300HS
  Detector Size (mm2)
315x315
315x315
300x300
300x300
  Xtal-Detector (mm)
110-900
130-800
100-850
100-900
  Throughput (images/min)
20
20
20
900
  Normal Rotation Method
Yes
Yes
Yes
Yes
  Grid Scan Search
No
No
No
Yes
  Helical Scan Method
No
No
No
Yes
  Multi-position Method
No
No
No
Yes
SAM
SAM
SAM
SAM
Yes
Yes
Yes
Yes